CÔNG TY TNHH ITVC TOÀN CẦU

Ngôn ngữ: vien

02253 261 208

Hours: 8AM-17PM (T2-T7)


Máy hiệu chuẩn đa năng

Máy hiệu chuẩn đa năng Fluke 741B 120

Máy hiệu chuẩn đa năng Fluke 741B 120

 
GiáVND 


Chi tiết sản phẩm

THÔNG SỐ KỸ THUẬT:
 

Measurement Accuracy

Voltage DC

110.000 mV

0.025%+0.015% *

1.10000 V

0.025%+0.005% *

11.0000 V

0.025%+0.005% *

110.000 V

0.05%+0.005% *

300.00 V

0.05%+0.005% *

Voltage AC

20 to 40 Hz

2% + 10

40 to 500 Hz

0.5% + 5

500 to 1 kHz

2% + 10

1 kHz to 5 kHz

10% + 20

Ranges:

1.1000, 11.000, 110.00, 300V

Current DC

30.000 mA

0.01% + 0.015% *

110.00 mA

0.01% + 0.015% *

Resistance

11.000 Ω

0.05% + 50 mΩ

11.000 Ω

0.05% + 50 mΩ

1.1000 kΩ

0.05% + 500 mΩ

11.000 kΩ

0.05% + 500 mΩ

Frequency

1.00 to 109.99 Hz

0.05 Hz

110.0 to 1099.9 Hz

0.5 Hz

1.100 to 10.999 kHz

5 Hz

11.00 to 50.00 kHz

50 Hz

Pressure

Accuracy

from 0.025% of range using any of 29 pressure modules. (for detailed specifications refer to pressure modules in options and accessories)
Modules available for differential, gauge, vacuum, absolute, dual and high pressure

Note

* (% of reading + % of full scale)

Source Accuracy

Voltage DC

110.000 mV

0.01%+0.005% *

1.10000 V

0.01%+0.005% *

15.0000 V

0.01%+0.005% *

Current DC

22.000 mA (Source)

0.01%+ 0.015% *

22.000 mA (Simulate)

0.02% + 0.03% *

Resistance

11.000 Ω
110.00 Ω
1.1000 kΩ
11.000 kΩ

0.01% + 20 mΩ
.01% + 40 mΩ
0.015% + 0.05 Ω
0.03% + 5Ω

Frequency

0.00 to 10.99 Hz

0.01 Hz

11.00 to 109.99 Hz

0.1 Hz

110.0 to 1099.9 Hz

0.1 Hz

1100 to 21999 Hz

2 Hz

22.000 to 50.000 kHz

5 Hz

Note

* (% of reading + % of full scale)

RTDs and Thermocouples

Measure and Source Accuracy:

10 Ω Cu (427):
100 Ω Pt (3916):
100 Ω Pt (3926):
100 Ω Pt (385):
200 Ω Pt (385):
500 Ω Pt (385):
1000 Ω Pt (385):
120 Ω Ni (672):
note:
E:
N:
J:
L:
K:
T:
U:
B:
R:
S:
C:
note:

2 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
For 2-and 3-wire measurement, add 0.4 °C
0.3 °C
0.5 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
0.9 °C
1.0 °C
0.9 °C
0.6 °C
Accuracy with external cold junction, for internal junction add 0.2°C

Source Accuracy:

10 Ω Cu (427):
100 Ω Pt (3916):
100 Ω Pt (3926):
100 Ω Pt (385):
200 Ω Pt (385):
500 Ω Pt (385):
1000 Ω Pt (385):
120 Ω Ni (672):
note:
E:
N:
J:
L:
K:
T:
U:
B:
R:
S:
C:
note:

1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
For 2-and 3-wire simulation, add 0.4 °C
0.2 °C
0.3 °C
0.2 °C
0.2 °C
0.3 °C
0.3 °C
0.3 °C
0.3 °C
0.8 °C
0.9 °C
0.9 °C
0.6 °C
Accuracy with external cold junction, for internal junction add 0.2 °C

Specifications

Ramp Functions

Source functions:

Voltage, current, resistance, frequency, temperature

Rate:

4 steps/second

Trip detect:

Continuity or voltage (continuity detection not available when sourcing current)

Loop Power Function

Voltage:

Selectable, 24 V or 28 V

Accuracy:

5%

Maximum current:

22 mA, short circuit protected

Maximum input voltage:

30 V DC

Step Functions

Source functions:

Voltage, current, resistance, frequency, temperature

Manual Step

Selectable step, change with arrow buttons

Autostep

Fully programmable for function, start delay, stepvalue, time per step, repeat

Environmental Specifications

Operating Temperature

-10 °C to +50 °C
-20 °C to +50 °C *

Storage Temperature

-20 °C to +60 °C

Dust/water resistance

Meets IP52, IEC 529

Operating Altitude

2800 m

Note

* Except frequency and AC

Safety Specifications

Agency Approvals

CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

Mechanical & General Specifications

Size

130 x 236 x 61 mm (5.1“ x 9.3” x 2.4”)

Weight

1.4 kg (3 lbs., 1oz.)

Batteries

NiCd: 7.2V, 1.7 Ah

Battery Life

~8 hours typical

Battery Replacement

Via snap-shut door without opening calibrator; no tools required

Side port connections

Pressure module connector
Connection for optional battery eliminator

Data storage capacity

1 day of calibration results

90 day specifications

The standard specification intervals for the 741B is 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "%of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.

Manufacturer : Fluke - USA

 Description

“A complete documenting calibrator”

The 741B does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. For documentation, the 741B automates calibration procedures and captures your data.

And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations. The 741B is the choice for plants that use standard company worksheets and that make reports in the shop. It has storage capacity for a day's calibration and measurement data. When you're back at the shop recall the data on screen to fill out calibration forms.

Features

The 741B is easy to use with flexible test strategies. And like all Fluke tools, it's rugged, reliable and tough as nails. It's like having several powerful tools in one - so you can work faster than ever:

  • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
  • Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
  • Power transmitters during test using loop supply with simultaneous mA measurement
  • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
  • Measure and source simultaneously with one compact, rugged, reliable tool
  • Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
  • Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc
  • Easy to use
  • Three year warranty
  • Bright white dual display. Read both sourced and measured parameters simultaneously.
  • Multi lingual interface
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
  • Includes
    • BC7217 battery charger
    • BP7217 battery pack
    • Instruction manual
    • NIST-traceable calibration report and data
    • One set of TP220 test probes
    • Two sets of AC220 test clips
    • Two sets of TL224 industrial test leads

Sản phẩm liên quan



Copyright © 2014 ICTV. All Rights Reserved.

tư vấn iso, tu van iso, kiểm toán năng lượng, kiem toan nang luong

02253 261 208