CÔNG TY TNHH ITVC TOÀN CẦU

Ngôn ngữ: vien

02253 261 208

Hours: 8AM-17PM (T2-T7)


ĐỒNG HỒ VẠN NĂNG

Đồng hồ vạn năng KYORITSU 1061, K1061 (True RMS)

Đồng hồ vạn năng KYORITSU 1061, K1061 (True RMS)

Đồng hồ vạn năng KYORITSU 1061, K1061 có các thông số kỹ thuật sau:

Chế độ : True RMS

DC V : 50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V

AC V : [RMS] 50.000/500.00mV/5.0000/50.000/500.00/1000.0V

DCV+ACV : 5.0000/50.000/500.00/1000.0V

DC A : 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A

AC A[RMS]  :500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A         

DCA+ACA  : 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A         

Ω : 500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ         

Continuity buzzer     : 500.0Ω     

Diode test : 2.4V                

C: 5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF

F: 2.000~9.999/9.00~99.99/90.0~999.9Hz/0.900~9.999/9.00~99.99kHz

Nhiệt độ : -200~1372ºC (K-type )         

Chức năng khác : Data Hold (D/H), Auto Hold (A/H), Peak Hold * (P/H), Range Hold (R/H), Maximum value (MAX), Minimum value (MIN), Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight         

Nguồn : R6P (1.5V)×4        

Kích thước : 192(L)×90(W)×49(D) mm         

Cân nặng     : 560g     

Nguồn :      LR44 (1.5V) × 2

Kích thước :192(L)×90(W)×49(D) mm

Khối lượng : 560g 

Phụ kiện : Vỏ, Pin, HDSD

Bảo hành : 12 tháng

Hãng sản xuất : Kyoritsu - Nhật

Xuất xứ : Nhật


Chi tiết sản phẩm

High Accuracy, High Performance and Reliable Measurements
0.02% basic DC accuracy
Large display with 50,000 counts
Dual display for double indication for AC and DC, V and Hz, etc.
True-RMS Measurements AC and AC+DC
Wide AC Frequency bandwidth from 10Hz to 100kHz *
True-RMS or MEAN value detection mode can be selected *
Fast Peak Hold response time of 250µs
Low-pass filter for motor drive measurements *
Low Power Ω for resistance measurements on sensible electronic circuits by a low and safe test current *
User calibration function
* Only for KEW 1062

1061/1062
CE
(Photo: 1062)
Safety design for industrial use
Complies with IEC 61010-1 CAT.III 1000V, CAT.IV 600V
Terminal shutter to prevent incorrect test leads is insertion in current terminals
Very wide operating temperature range from -20 to 55ºC
Reliable support for data management
Large data internal memory
  KEW 1062: 10,000 data (logging), 100 data (manual)
  KEW 1061: 1,000 data (logging), 100 data (manual)
Download data and Live Monitoring on a PC via the USB interface
(Option for USB Communication set)

TRUE RMS CAT.IV 600V
 
KEW 1061 KEW 1062
Detection mode RMS MEAN/RMS (switch)
DC V 50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V
(Input impedance: Approx. 100MΩ [50/500/2400mV], 10MΩ [5/50/500/1000V])

|±0.02%rdg±2dgt (Basic accuracy)
AC V
[RMS]
50.000/500.00mV/5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V])
|±0.7%rdg±30dgt (Basic accuracy) |±0.4%rdg±30dgt (Basic accuracy)
AC V
[MEAN] *
-- 50.000/500.00mV/ 5.0000/50.000/500.00/1000.0V
(Input impedance:
11MΩ<50pF [50/500mV/5V],
10MΩ<50pF [50/500/1000V])


|±1%rdg±30dgt (Basic accuracy)
DCV+ACV 5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [5V], 10MΩ<50pF [50/500/1000V])
|±1%rdg±10dgt (Basic accuracy) |±0.5%rdg±10dgt (Basic accuracy)
DC A 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A

|±0.2%rdg±5dgt (Basic accuracy)
AC A
[RMS]
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±1%rdg±20dgt (Basic accuracy) |±0.75%rdg±20dgt
|(Basic accuracy)
AC A
[MEAN] *
-- 500.00/5000.0µA/50.000/500.00mA/ 5.0000/10.000A

|±1.5%rdg±20dgt (Basic accuracy)
DCA+ACA 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±1.5%rdg±10dgt (Basic accuracy) |±1%rdg±10dgt (Basic accuracy)
Ω 500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ